394C06/M394C06加速度标定器快速报价
Platinum Stock Product Handheld shaker, 1g at 159.2 Hz (for upto 210 grams total weight of sensor, cable and mtg.accessories).
Acceleration Output: (±3%) 1.00 g rms(9.81 m/s² rms)
Velocity Output: 0.39 in/sec rms(9.81 mm/s rms)
Operating Frequency: (±1%) 159.2 Hz
Maximum Load: 7.4 oz(210 gm)
Weight: (with batteries) 31 oz(900 gm)
English:SI: PERFORMANCEOperating Frequency (±1%)159.2 Hz159.2 Hz
Acceleration Output (±3 %)1.00 g rms9.81 m/s² rms[5] VelocityOutput0.39 in/sec rms9.81 mm/s rms[4] Displacement Output0.39mil rms9.81 µm rms[4] Transverse Output≤3 %≤3 %
Distortion (0 to 99 grams load)≤2 %≤2 %
Distortion (100 to 149 grams load)≤5 %≤5 %
Distortion (150 to 210 grams load)≤9 %≤9 %
Maximum Load7.4 oz210 gm[1] Automatic Switch Off Time1.0 to2.5 minutes1.0 to 2.5 minutes[3] Calibration Cycles (2 gramload)320 cycles320 cycles[2] Calibration Cycles (25 gramload)600 cycles600 cycles[2] Calibration Cycles (50 gramload)1600 cycles1600 cycles[2] Calibration Cycles (100 gramload)400 cycles400 cycles[2] Calibration Cycles (150 gramload)160 cycles160 cycles[2] Calibration Cycles (210 gramload)80 cycles80 cycles[2] ENVIRONMENTALTemperature Range(Operating)+15 to +130 °F-10 to +55 °C
ELECTRICALRamp-Up time≤3 sec≤3 sec[6] Internal Battery(Quantity)44
Internal Battery (Type)AAAA[7] DC Power (±5 %)12 VDC12VDC[9] DC Power1.67 Amps1.67 Amps[8][9] Battery Life (2gram load)8 hours8 hours[2] Battery Life (25 gram load)15hours15 hours[2] Battery Life (50 gram load)40 hours40hours[2] Battery Life (100 gram load)10 hours10hours[2] Battery Life (150 gram load)4 hours4hours[2] Battery Life (210 gram load)2 hours2hours[2] PHYSICALSize - Diameter2.2 in56 mm
Size - Height7.8 in200 mm
Weight (with batteries)31 oz900 gm[6] Mounting Thread10-32Female10-32 Female[10][11] Mounting Torque (Maximum)10in-lb112 N-cm[11]
All specifications are at roomtemperature unless otherwise specified.
Product Notes
[1] Maximum load includes sensor, connector and cabling.
[2] Approximate values, based on automatic switch off time anddependent on type of batteries.
[3] Unit supplied set to auto shut off; see manual for continuoususe selection.
[4] Calculated values for reference only.
[5] Unit supplied set to rms; see manual for peak selection.
[6] Typical.
[7] Alkaline type recommended for longest service life.
[8] Maximum.
[9] This specification for external DC power supply (optional).
[10] Test sensor should be hand tightened (without tools).
[11] Transducer to shaker table.
[12] See PCB Declaration of Conformance PS022 for details.
Accessories:
Optional073A16
(1)
080A150Mounting Base (1/4-28)(1)
080B443-Pin Mounting Adapter(1)
Supplied073A15Battery Pack(1)
080A109Petro Wax(1)
080A178Adhesive Mounting Base with 1/2" hex 10-32 IntegralStud(1)
080A84Mounting Base (5-40 to 10-32)(1)
080A85Mounting Base (M3 X 0.5 to 10-32)(1)
081A08Mounting Stud (10-32 to 1/4-28)(1)
081B05Mounting Stud (10-32 to 10-32)(2)
M081B05Mounting Stud 10-32 to M6 X 0.75(1)
M081B23Metric mounting stud, 10-32 to M5 x 0.80 long(1)
Optional Versions:
(M) Metric Mount
ENGLISHSI Acceleration Output (±3 %)1.02 grms10.0 m/s² rms